FDAClass II · Medical device · Recalled July 8, 2026
Short Monobloc Stem Broaches recall — Medacta Usa Inc
Hazard: Surgical instruments used during shoulder arthroplasty procedures does not progressively translate to replicate implant geometry, so the reaming guide and the stem reamer are positioned approximately 0.2 mm per size more laterally than intended, which may increase the risk of fracture in the medial region of the proximal humerus.
Class II: may cause temporary or medically reversible health consequences.
View the official FDA enforcement record (openFDA) ↗
Description
Short Monobloc Stem Broaches, Size/REF: 7/04.01.10.0970, 8/04.01.10.0971, 9/04.01.10.0972, 10/04.01.10.0973, 11/04.01.10.0974, 12/04.01.10.0975, 13/04.01.10.0976, 14/04.01.10.0977, 15/04.01.10.0978, 16/04.01.10.0979 Part of Medacta Shoulder System
Affected products
Companies & origin